Dear Marc

We recommend cleaning dispensing tips with Hellmanex II from the German
company Hellma.  It's for cleaning cuvettes.  You can buy it from VWR
and others.

Hellmanex is a mildly alkaline solution with surfactants etc.  It has no
enzymes in it, but you do need to flush with buffer to get rid of the
alkali.

Previously, users reported that cleaning with methanol mixed with
concentrated HCl worked very well in extreme cases.

Our robot, the Oryx, uses only one tip for protein.  This is
semi-disposable.  Using one tip and touching off the drops has the great
advantage that virtually no protein is wasted, and drops as low as 20 nl
can be dispensed, even containing 50% glycerol!

We recommend that users keep tips that have been used for proteases
separately, and clean them after use.  (The system comes with 8 tips and
replacements cost 45 USD.)  Also we recommend that tips are thoroughly
cleaned once a fortnight for average use in any case.

I hope this is helpful.

Sincerely

Patrick


--  
[EMAIL PROTECTED]    Douglas Instruments Ltd.  
DouglasHouse, EastGarston, Hungerford, Berkshire, RG177HD, UK
Directors: Peter Baldock, Patrick Shaw Stewart, James Smith
http://douglas.co.uk or http://www.douglasinstruments.com
Tel: 44 (0) 148-864-9090    US toll-free 1-877-225-2034

> -----Original Message-----
> From: [EMAIL PROTECTED] [mailto:[EMAIL PROTECTED] On Behalf
Of Marc
> Graille
> Sent: 19 December 2005 12:30
> To: Patrick Shaw Stewart
> Subject: [ccp4bb]: Crystallization robots and protease.
> 
> ***  For details on how to be removed from this list visit the  ***
> ***          CCP4 home page http://www.ccp4.ac.uk         ***
> 
> 
> Dear all,
> 
> I have a question regarding the use of robotics to screen for
> crystallization conditions for proteases.
> Does anyone have already used robots on proteases? If yes, have you
> experienced any protease "contaminant" in the robot pipes, which could
> have affected the results on other projects performed during the next
> few days ??
> I mean that we cannot exclude that a "contaminant" could digest the
> protein we are working on and yield crystals of a fragment of the
> studied protein.
> 
> We are hesitating in using our crystallization robots on proteases as
we
> are afraid to have some contaminant in the pipes that will disturb all
> our future experiments!!!
> Any advice about how to clean the robot syringes after use of
proteases
> are welcome!!!
> 
> 
> Regards,
> 
> Marc
> 
> --