Have all crystals ready (pre-mounted in canes) and transfer them quickly !
We make it a three-person job, one takes notes, one removes the vail
from the canes and dumps it into a small bowl containing LN2 and the
third person trnasfers the crystal into the cassette. I personally don't
use their mounting device but the Hampton magnetic wand with eject
button - for me this works much faster (and I will soon be flammed by
the SSRL guys for doing so).
Have enough LN2 ready to refill the blue foam container and keep
refilling it. A full 96 cassette should not take more than 20 minutes to
fill up in this setting. When you are used to it you can do it in 15
minutes.
Juergen
Chunmin L wrote:
*** For details on how to be removed from this list visit the ***
*** CCP4 home page http://www.ccp4.ac.uk ***
Hi,
We are preparing a remote data collection at SSRL. Last time, we had
icing problem when we load mounted crystals into the sample cassette.
Any tips to reduce or prevent it are very welcome.
Thanks in advance!
Chunmin
--