Hi Ian,

I am happy to read that your users trust d*TREK so much that they simply click the buttons and away they go. The indexing program within d*TREK now outputs the following text:

       INFO - The above indexing solution is ONLY a hypothesis.
       One must confirm the hypothesis by examining symmetry
       of observed intensities as well as possible systematic
       absences.
       One does this by collecting data, processing the data
       and using dtcell and dtscaleaverage.

But my guess is that folks don't read the output in this click-happy world. Indeed, as I have told many folks, the indexing step is the first step in a crystal structure determination where one might need to know some crystallography before proceeding. One can clone, express, purify, ..., grow crystals, harvest, flash-cool them, take diffraction images -- all without even knowing that there was someone named Bragg who had a Law.

There are even scripts or pipelines where users put in multiple spacegroups and let the algorithm decide which one works to solve the structure.

So I view this as one of education. It is probably easier to teach your colleagues how to transform cell axes than it is to teach them how to interpret or even write Hermann-Mauguin symbols.

d*TREK can help you with the transformation. One can transform cell axes and reindex diffraction datasets trivially with dtcell.

One can even force dtindex to only look for a=70, b=90, c=60 and spacegroup 18 as P2(1)2(1)2, if you like. So I cannot imagine that your problems cannot be solved with existing solutions. First, confirm with the user that the spacegroup number is 18 and that they want the 60 Angstrom axis to not have systematic absences. Then you can use
  dtindex ... -cell 70 90 60 90 90 90 -known -spacegroup 18 ...
This can be used in scripts or in the GUI. For more help on dtcell, see "dtcell -h", but a possible option is
  dtcell  ... -transform bca ...

Cheers,

Jim

Subject: [ccp4bb] Can't choose full H-M symbol for conventional cell ?


We're having some problems essentially because AFAICS Mosflm (and also
D*Trek) don't recognise the 'full' Hermann-Mauguin symbols for space
groups.

For example for space group no 18 with the 'short' H-M symbol 'P 2/1 2/1
2' in the setting with 'full' H-M symbol 'P 2 2/1 2/1' the programs
choose the correct conventional cell as say a=60, b=70, c=90 (according
to the convention a<=b<=c).  But then unless I've missed something
...

Reply via email to