Vin,

This effect is what you see when the goniometer's rotation axis is not exactly perpendicular to the beam. HKL2000 is compensating for rotations that are actually due to goniometer misalignment by changing the crystal orientation angles. It does not cause any harm to your data, unless things are so far off that refinement gets lost, which does not seem to have been the case for your data.

Marian Szebenyi
MacCHESS

Vin Purp wrote:
Hi, sorry for the non-ccp4 related question.
I have one data set that yields peculiar results when I use HKL2000 (or
3000) to integrate it.
Background: I have crystals that index to the I222 sg. I have shot and
collected datasets, integrated, molreped and refined  many hundreds of these
same crystals and know them very well. However, I recently collected data
that has the following 'problem' when I integrate: I collected 450 frames at
0.3 width (135 deg) .The Rot. Change vs. Frame window in the HKL program
(Integration tab) shows that in the first 150 frames, the x and z  deviates
+30 and frames 150-450 x and z cross to a negative change and finish at
-120. The Y-axis doesn't deviate from 0 the entire time.
All other statistics look 'normal' for the integration; mosaicity is a bit
high (0.75-1), the Chi sq. hovers near 1 for all frames, and the 'Cell' and
'Distance' vs. frame do not deviate from 0 much at all over all frames. I
have played with integration box size, spot size, elongation and background
with the same results.

Other stats: data set is ~1.60 Ang. resolution; unit cell = 128x140x168
Ang.; Roverall = 4.2%
 I was curious as to what this phenomena is due to. Any ideas are appreciated.