Hi Experts,

Last year I collected a dataset at 6 A of 2600 aa multi-domain protein. 

Resolution range 50 - 6 (6.7-6)
Wilson B    370 A^2
Reflection 34120 (12240)
Rmeas 0.2 (4.4)
Rpim 0.06(1.2)
I-sigI  9 (1)
CC1/2   0.9 (0.5)
 
I have the following questions in my mind.

1. Does it make any sense to solve the structure at _this_ resolution? It is 
not completely novel protein, there are known structures with about 54 % 
identity to this. The fold is known to be the same. 

2. Doing a molecular replacement with "Phaser" using EM-model gives me a unique 
solution. And I can see a reasonable electron density map. Tried again with 
AMORE - The amore-build-output model also gives the same solution. In both 
cases, the solutions are unique. There are no so-called translational symmetry.

3. If I do a Refmac restrained refinement, though I get R/Rfree in around 
30.1/35.5 the stereochemistry is very poor (18 % outliers). I had to enable 
tight WEIGHT MATRIX (1e-7). Here at this resolution does it make any sense to a 
restrained refinement?

4. If I do only a rigid body with Refmac, the R-factor/Rfree are at around 41 
%, and in many places model does not fit density. When I manually correct these 
and refine there is basically no change in R-factor/R-free (it even worsens in 
cases).

>From: Bernhard Rupp (Hofkristallrat a.D.) <[email protected]>>Date: 27 
>April 2015 at 21:54
>Subject: Re: [ccp4bb] 3BDN, 16.5% Ramachandran Outliers!!!!!
>To: [email protected]
>What we cannot tell sans supporting density is whether it is a more accurate 
>model, although I have>rarely seen an improvement in geometry giving worse 
>density fit. Usually a mess remains a mess - 
>there is (at this resolution) no free lunch. The key question is again – does 
>the model justify 

5. I did try DEN, reference model restraints, PROSMART(refmac) etc which have 
no improvement at all - R/Rfree stuck at 41 %. 

6. Since my group is EM-group, I wonder when EM-maps of 6 A are published, why 
are X-ray data at the same resolution not being published? What happens to 
these datasets?

7. Can I _just_ do a molecular replacement and just mutate residues (based on a 
sequence alignment - There are large numbers of deletions and hence sequence 
registers are different/unknown) and deposit it as a model in the PDB? Should I 
put the side-chains or it is meaning less at this resolution? Why in the 
EM-field they are allowed to deposit such coordinates with side-chains?

8. As [email protected] points out 

>Particularly in Molecular Replacement structures, and here particularly in 
>those with multi->segment/domain models, there are almost always parts that 
>fit well and others>that fit poorly -  with simply not enough data at the 
>given resolution to improve the poor parts>sans additional phase information. 
>Bias issues have been discussed and need not be iterated here.
Since my protein also has multiple catalytic domains, some of them better 
resolved and others terrible. What about bias here at 6 A resolution? Very 
large problem, indeed.

Apologies for the long email, and any suggestion will be gratefully received.

Many thanks,
Kerstin



---------- Forwarded message ----------
From: Bernhard Rupp (Hofkristallrat a.D.) <[email protected]>
Date: 27 April 2015 at 21:54
Subject: Re: [ccp4bb] 3BDN, 16.5% Ramachandran Outliers!!!!!
To: [email protected]

I’d be very careful at judging low resolution structures. This is a tricky 
businessrequiring a lot more info than just the PDB validation report. The 3+ 
to 4 Aresolution range is a particularly deceptive one: The crystallographer 
does not have much data given the model parameters (perhaps consulting his 
figure showingdeterminacy for coordinate refinement might 
help)http://www.ruppweb.org/Garland/gallery/Ch12/pages/Biomolecular_Crystallography_Fig_12-11.htm
 At this resolution one has about enough data to keep enthusiasm up but at the 
same time it isnot quite yet bad enough to throw up the hands and admit that 
that one is de facto modelling with a few X-ray restraints (i.e. data), 
requiring correspondingly suitable refinement protocols (and discipline,aka 
mental restraints in addition to stereochemical restraints). One is easily 
spoiled by looking exceptional 2A structures of huge complexes, butnature (I do 
not mean the journal but the same time would not exclude it) is often cruel. 
Particularly in Molecular Replacement structures, and here particularly in 
those with multi-segment/domain models, there are almost always parts that fit 
well and othersthat fit poorly -  with simply not enough data at the given 
resolution to improve the poor parts sans additional phase information. Bias 
issues have been discussed and need not be iterated here. Pavel is correct in 
pointing out that a model with better geometry is also a more plausible 
model.What we cannot tell sans supporting density is whether it is a more 
accurate model, although I haverarely seen an improvement in geometry giving 
worse density fit. Usually a mess remains a mess - there is (at this 
resolution) no free lunch. The key question is again – does the model justify 
the specific conclusions drawn from it? If a poor model is better than no model 
at all, be it, as longas this is recognized and not used as an excuse for 
careless work. Facile dictu, difficile factu.
                                          