Hi Paul,

It's just a matter of adjusting the SensorCal menu setting until the power reads the same as the device used for the calibration. Of course, that assumes the calibration device is more accurate than the P3. :=)

The nominal calibration value is 500. The intent was to make it correspond to the same value as in a W2, but the match is not perfect. For one thing, the P3 uses a more-sophisticated algorithm to match the diode detector nonlinearity than the W2.

Alan N1AL


On 03/06/2017 08:56 AM, paul ecker via Elecraft wrote:
Alan - is there any documentation available on the procedure for calibrating 
the Tx Monitor?
73Paul