When I tried setting K3 CONFIG:SMTR MD to ABS, I found that the S-meter reading then stayed constant regardless of the setting of the preamp/attenuator, in agreement with the Owner's Manual addendum.
However, the reading for a 50-microvolt input signal (with the preamp on and attenuator off) falls from S-9 to S-6 as CONFIG:SMTR MD is changed from NOR to ABS. Shouldn't a 50-microvolt input signal produce an S-9 reading regardless of the setting of CONFIG:SMTR MD? Alternatively, the Owner's Manual S-Meter calibration procedure would need to mention the setting of CONFIG:SMTR MD. 73, Paul W8TM _______________________________________________ Elecraft mailing list Post to: [email protected] You must be a subscriber to post to the list. Subscriber Info (Addr. Change, sub, unsub etc.): http://mailman.qth.net/mailman/listinfo/elecraft Help: http://mailman.qth.net/subscribers.htm Elecraft web page: http://www.elecraft.com