Allen,

Using a room within the company for radiated measurement is not such a good
idea. Many engineers including myself attemted this, but had no luck as
there are too many steels and metal work  used in the construction of the
building. You will not even come close to the test house as far as the
repeatability is concerned. I suggest you ask the VP of your opreration to
allocate you part of the carpark as well as this room. You can use the room
for conducted emissions and the carpark for radiated. That is if you are
not close to any cab office, fire station, or heavy industrial site.

Good luck