this might have to become an offline topic.
 
I am interested in what other labs are doing with regard to LAN/ethernet ports
on ITE products sold as Medical equipment. 
 
60601 calls up CISPR11, which doesn't have Signal Port conducted emission
requirements as CISPR22 does.
 
In this age of litigation I am interested if any labs are testing sig port
under medical emc specs and reporting it as a deviation ?
 
Appreciate any feedback via forum or off line.
 
rgds
 
Julian
 
 
 
 
 
- ---------------------------------------------------------------- This
message is from the IEEE Product Safety Engineering Society emc-pstc
discussion list. Website: http://www.ieee-pses.org/ 

To post a message to the list, send your e-mail to [email protected] 

Instructions: http://listserv.ieee.org/request/user-guide.html 

List rules: http://www.ieee-pses.org/listrules.html 

For help, send mail to the list administrators: 

Scott Douglas [email protected] Mike Cantwell [email protected] 

For policy questions, send mail to: 

Jim Bacher: [email protected] David Heald: [email protected] 

All emc-pstc postings are archived and searchable on the web at: 

http://www.ieeecommunities.org/emc-pstc 

Reply via email to