this might have to become an offline topic. I am interested in what other labs are doing with regard to LAN/ethernet ports on ITE products sold as Medical equipment. 60601 calls up CISPR11, which doesn't have Signal Port conducted emission requirements as CISPR22 does. In this age of litigation I am interested if any labs are testing sig port under medical emc specs and reporting it as a deviation ? Appreciate any feedback via forum or off line. rgds Julian - ---------------------------------------------------------------- This message is from the IEEE Product Safety Engineering Society emc-pstc discussion list. Website: http://www.ieee-pses.org/
To post a message to the list, send your e-mail to [email protected] Instructions: http://listserv.ieee.org/request/user-guide.html List rules: http://www.ieee-pses.org/listrules.html For help, send mail to the list administrators: Scott Douglas [email protected] Mike Cantwell [email protected] For policy questions, send mail to: Jim Bacher: [email protected] David Heald: [email protected] All emc-pstc postings are archived and searchable on the web at: http://www.ieeecommunities.org/emc-pstc