On Wednesday, February 20, 2013 04:41:31 PM Bruce Ravel wrote: > > but understand your concerns about the data I sent as they were just > > test files we collected when we were setting up the beam. However, I > > do not think the abridged range is the cause of the problem, as I > > wasl having issues importing real data as well (i.e. -150 eV > > pre-edge and k=12 post edge). These spectra files are attached, > > along with the error log file and the dathena.log file that was > > recorded when importing these files. > > You are correct. When importing multiple data sets through the HXMA > filetype plugin, the reference channels for all but the first file are > imported incorrectly. I am not yet sure what is causing the problem, > but I will look into it and try to have a solution in place for the > next release. > > Unfortunately, the only workaround entirely within Athena is to import > the data one at a time.
As I look into this problem, and with some additional useful information from Fred Mosselmans, I see that this is a pervasive problem (and not specific to the HXMA plugin) that had somehow escaped my attention. Yikes! I have been working on this today. Hopefully, I will have a solution soon. B -- Bruce Ravel ------------------------------------ bra...@bnl.gov National Institute of Standards and Technology Synchrotron Methods Group at NSLS --- Beamlines U7A, X24A, X23A2 Building 535A Upton NY, 11973 Homepage: http://xafs.org/BruceRavel Software: https://github.com/bruceravel _______________________________________________ Ifeffit mailing list Ifeffit@millenia.cars.aps.anl.gov http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit