Hello Yanny:

I've modeled a few Pt EXAFS spectra and have found that using an E0 at the top 
of the white line usually does a better job of getting the data close to the 
theory. This seems to be often the case with the L3-edges as they are rather 
broad compared to the K-edges of transition metals.  If you would like to share 
your project files with me. I can take a look.  14 eV seems to be too much and 
maybe an indication that there is a different type of atom in the first shell.

B.T.W. Using the first derivative is great for XANES analysis, but often not 
the best choice for EXAFS.

Kind regards,
Shelly Kelly (ske...@anl.gov)

From: Ifeffit <ifeffit-boun...@millenia.cars.aps.anl.gov> On Behalf Of Zi Qi 
Chen
Sent: Tuesday, August 22, 2023 9:53 AM
To: ifeffit@millenia.cars.aps.anl.gov
Subject: [Ifeffit] Shift in Eo when fitting PtO2 and similar samples

Dear all,

I am looking at EXAFS of PtO2 powders and Pt particles supported on TiO2. I 
used a beta-PtO2 crystal structure data for the PtO2 EXAFS and after the first 
shell fit, I observed a large shift in Eo (delta Eo of ~14 eV) in my fit 
parameters. When I tried to fit the Pt foil to obtain the So2, the energy shift 
also seemed high (delta Eo of ~7 eV). I am wondering if there could be a 
problem in the selection of Eo prior to EXAFS analysis? I left the Eo as the 
energy at the first peak of first derivative (ifeffit default).
All of the data were also aligned using the reference foil in Athena before 
looking at EXAFS. I am wondering what would have caused this shift? Any help is 
appreciated, thank you.

Best regards,
Yanny