Joe, Monticello Nuclear Generating Plant uses three of these units.
Our processes are still evolving. Our biggest learning curve has been on the use of QC charts for daily function checks versus the daily determination of background and efficiency. It is a new process for most of the technicians. Our other issue is the accumulation and how much data to acquire to generate the QC charts. This has caused problems with having daily function checks outside the 2SD or 3SD sometimes because our tolerances are too tight from limited data with small variation; conversely, we gather data for 6 weeks which greatly extends the time needed for daily checks because we are taking multiple counts per test. The last issue has to do with making sure the technicians know the correct responses for when the daily checks are outside the 2SD or 3SD limits. Hope this is useful, Scott Huneycutt Xcel Energy | Responsible By Nature Senior Health Physicist Monticello Nuclear Generating Plant 2807 West County Road 75, Monticello, MN 55362 P: 763.295.1380 F: 763.295.1225 E: [email protected]<mailto:[email protected]> ________________________________________________ <blocked::blocked::http://www.xcelenergy.com/>XCELENERGY.COM<http://www.xcelenergy.com/> Please consider the environment before printing this email ________________________________ From: [email protected] [mailto:[email protected]] On Behalf Of [email protected] Sent: Thursday, June 09, 2011 6:54 AM To: [email protected] Subject: Powernet: Canberra alpha/beta counter Cook Plant would like to know if anyone is using a Canberra Series 5 XLB gas flow proportional alpha/beta counting system with the Eclipse software. Please provide a contact to discuss your operating experience with this instrument. Thank you. Joe Beer, CHP Staff Health Physicist Cook Nuclear Plant, Mail Zone 9 One Cook Place Bridgman, MI 49106 audinet: 280-2262 work: 269-465-5901 ext 2262 fax: 269-466-2661 cell: 269-313-7782 [email protected] ---------------------------------------------------------------- This e-mail message and all attachments transmitted with it from the Nuclear Generation Group of American Electric Power are for the sole use of the intended recipient(s) and may contain confidential and privileged information. Any unauthorized review, use, disclosure or distribution is prohibited. If you are not the intended recipient, please contact the sender by reply e-mail and destroy all copies of the original message.