what is the justification for using half that? 7.3 is not high in
general. what scanner are you using?
On 5/8/2023 4:27 AM, Federica Di Antonio wrote:
External Email - Use Caution
Dear experts,
I am here again because I'd like to ask a
External Email - Use Caution
Dear experts,
I am here again because I'd like to ask a specification about the fwhm
that the alghoritm evaluates: the main value that I obtained was 7.3 mm; It
is an high value for a pv correction.
May I put in half the value?
Because I wonder if the
External Email - Use Caution
Thank you so much!
Il giorno gio 15 dic 2022 alle ore 16:24 Douglas N. Greve <
dgr...@mgh.harvard.edu> ha scritto:
> I have something called the adaptive GTM that is currently a "hidden"
> feature in mri_gtmpvc as it has not been published yet. You
I have something called the adaptive GTM that is currently a "hidden"
feature in mri_gtmpvc as it has not been published yet. You can try it
by adding the following flags --opt 1 --opt-tol 4 10e-6 .02 --opt-seg-merge
When this is done, there will be a file called gtmdir/aux/opt.params.dat
The
External Email - Use Caution
Dear experts,
Few weeks ago i wrote to this mailing list but I didn't receive any answer,
so I'll try again:
mri_gtmpvc --i pet.nii.gz --reg template.reg.lta --psf FWHM --seg gtmseg.mgz
--default-seg-merge --auto-mask 1 .01 --mgx .01 --o
External Email - Use CautionDear experts, mri_gtmpvc --i pet.nii.gz --reg template.reg.lta --psf FWHM --seg gtmseg.mgz --default-seg-merge --auto-mask 1 .01 --mgx .01 --o gtmpvc.output About this command, is there the possibility for PETsurfer to compute the fwhm of the psf of