Hi Ben,
Yes, the CMRR multiband sequence incorporates "dynamic off-resonance in
k-space” (DORK)” method as in Pfeuffer et al., MRM 2002, which has already been
implemented in the Siemens product EPI sequence ICE recon.
Gordon
> On Jun 8, 2017, at 12:05 PM, Benjamin Risk
Harms,
> Michael" <mha...@wustl.edu>
> Date: Thursday, June 8, 2017 at 2:05 PM
> To: Benjamin Risk <bb...@cornell.edu>, "hcp-users@humanconnectome.org" <
> hcp-users@humanconnectome.org>
> Cc: Essa Yacoub <yaco0...@umn.edu>
> Subject: Re: [HCP-Us
org<mailto:hcp-users@humanconnectome.org>"
<hcp-users@humanconnectome.org<mailto:hcp-users@humanconnectome.org>>
Subject: [HCP-Users] Question about unprocessed fMRI data
Dear HCP experts,
I am trying to better understand the unprocessed fMRI data. Does the online
re
enjamin Risk <bb...@cornell.edu<mailto:bb...@cornell.edu>>
Date: Thursday, June 8, 2017 at 11:05 AM
To: "hcp-users@humanconnectome.org<mailto:hcp-users@humanconnectome.org>"
<hcp-users@humanconnectome.org<mailto:hcp-users@humanconnectome.org>>
Subject: [HCP-U
Dear HCP experts,
I am trying to better understand the unprocessed fMRI data. Does the online
reconstruction include corrections for spatial shifts due to B0
fluctuations? There is a note about this in Ugurbil et al 2013 but I am
wondering if this applies to the final HCP acquisition protocol.