[Ifeffit] problem with feeff.inp

2008-02-21 Thread Sebastiano Cammelli
Dear
I have some problem with a very easy simulation. Everytime I try to introduce 
a potential with more elements, for instance

POTENTIALS
0 26 Fe
1 26 Fe 
2 29 Cu

the program stops the simulation with a message that I do not understand and 
that is attached to this mail. This 'weird' behavior does not happen with a 
monoatomic potential like 

POTENTIALS
0 26 Fe
1 26 Fe 

and in this second case everything works perfectly.
I would be very glad if you will have any suggestion. 
Thank you in advance
Sebastiano Cammelli 
feff.inp1: this works well

TITLE Fe 
CONTROL 1 1 1 1 1 1
PRINT 0 0 0 0 0 0
LDOS -30 20 0.2 
EDGE K
S02 1.0
SCF 4 0 30 0.2 1 
XANES 4 0.07 0.0
FMS 6 0   
POTENTIALS
0 26 Fe
1 26 Fe 
ATOMS
  0.00.00.00 Fe   0.0
1.433251.433251.43325  1 Fe_12.48246
   -1.433251.433251.43325  1 Fe_12.48246
1.43325   -1.433251.43325  1 Fe_12.48246
   -1.43325   -1.433251.43325  1 Fe_12.48246
1.433251.43325   -1.43325  1 Fe_12.48246
   -1.433251.43325   -1.43325  1 Fe_12.48246
1.43325   -1.43325   -1.43325  1 Fe_12.48246
   -1.43325   -1.43325   -1.43325  1 Fe_12.48246
*...
END

feff.inp2: this does not work, the only difference is the atom of Cu in first 
shell. 

TITLE Fe 
CONTROL 1 1 1 1 1 1
PRINT 0 0 0 0 0 0
LDOS -30 20 0.2 
EDGE K
S02 1.0
SCF 4 0 30 0.2 1 
XANES 4 0.07 0.0
FMS 6 0   
POTENTIALS
0 26 Fe
1 26 Fe 
2 29 Cu 
ATOMS
  0.00.00.00 Fe   0.0
1.433251.433251.43325  1 Fe_12.48246
   -1.433251.433251.43325  1 Fe_12.48246
1.43325   -1.433251.43325  1 Fe_12.48246
   -1.43325   -1.433251.43325  2 Cu_12.48246
1.433251.43325   -1.43325  1 Fe_12.48246
   -1.433251.43325   -1.43325  1 Fe_12.48246
1.43325   -1.43325   -1.43325  1 Fe_12.48246
   -1.43325   -1.43325   -1.43325  1 Fe_12.48246
*
END

in this second case the message is: 

[EMAIL PROTECTED] 2]$ ./feff
 Feff 8.40
  XANES:
 Fe in Fe 7
 Calculating potentials ...
free atom potential and density for atom type0
free atom potential and density for atom type1
free atom potential and density for atom type2
initial state energy
overlapped potential and density for unique potential0
overlapped potential and density for unique potential1
overlapped potential and density for unique potential2
muffin tin radii and interstitial parameters
 iph, rnrm(iph)*bohr, rmt(iph)*bohr, folp(iph)
0  1.40339E+00  1.35427E+00  1.09245E+00
1  1.41347E+00  1.36196E+00  1.09680E+00
2  1.27400E+00  1.24500E+00  1.05748E+00
  Core-valence separation
 mu_old=   -11.483
  SCF ITERATION NUMBER  1  OUT OF 30
 Calculating energy and space dependent l-DOS.
 It takes time ...
 point #   1  energy = -40.000
Doing FMS for a cluster of  15 atoms around iph =  0
0   FMS matrix (LUD) at point   1, number of state kets = 135
Doing FMS for a cluster of  15 atoms around iph =  1
0   FMS matrix (LUD) at point   1, number of state kets = 135
Doing FMS for a cluster of  15 atoms around iph =  2
0   FMS matrix (LUD) at point   1, number of state kets = 135
 point #  20  energy = -28.815
 point #  40  energy = -16.459
 point #  60  energy = -15.827
  Electronic configuration
   iphil  N_el
 0 00.669
 0 10.939
 0 28.180
 0 30.000
 1 00.670
 1 10.946
 1 26.522
 1 30.000
 2 00.430
 2 10.522
 2 20.557
 Found bad counts.
  Occupation number in getorb is10.000
  Will repeat this iteration
 2 30.000
./feff: line 4: 12214 Segmentation fault  $FeffPath/ffmod1
 Calculating LDOS ...
open: No such file or directory
apparent state: unit 3 named pot.bin
lately writing sequential formatted external IO
./feff: line 5: 12217 Aborted $FeffPath/ldos
 Calculating cross-section and phases...
open: No such file or directory
apparent state: unit 3 named pot.bin
lately writing sequential formatted external IO
./feff: line 7: 12218 Aborted $FeffPath/ffmod2
 Error opening file, phase.bin in module rdxsph
 Fatal error
CHOPEN
STOP   statement executed
 Preparing plane wave scattering amplitudes...
 Error opening file, phase.bin in module rdxsph
 Fatal error
CHOPEN
STOP   statement executed
 Calculating EXAFS parameters...
 Error opening file, phase.bin in module rdxsph
 Fatal error
CHOPEN
STOP   statement executed
 Calculating chi...
fmt: end of file
apparent state: unit 8 named fort.8
last format: (A)
lately reading sequential formatted external IO
./feff: line 11: 

Re: [Ifeffit] Questions about Athena and XANES

2008-02-21 Thread Jeff Terry

Hi Jenny,

I just wanted to add a bit to Matt's answer on TEY vs FY and suggest a  
couple of reference for you.


There are three contributions to the TEY: photoelectrons, Auger  
Electrons, and secondary electrons. The secondary electrons will  
typically determine the maximum depth that you can observe in a TEY  
measurement. There is a good description of this in Jo Stohr's book,  
NEXAFS Spectroscopy which can be seen at google books.

http://tiny.cc/44VGq

There is also a quick synopsis of the book at: 
http://www-ssrl.slac.stanford.edu/stohr/nexafs.htm
At this site, there is a photoelectron spectrum showing the relative  
intensity of Auger and Photopeaks. The secondaries are roughly a  
factor of 100 more intense than the other contributions.


It is possible to measure both TEY and Auger electron yield (AEY) and  
you will see that the AEY is more surface sensitive that TEY.  
Measuring AEY requires an analyzer capable of discriminating the  
electron energy. AEY is typically performed at lower photon energies  
as there are not too many electron energy analyzers at hard x-ray  
facilities.


It is actually pretty difficult to try to calculate the maximum  
analysis depth for TEY because of the three different contributions.  
Here is an attempt to understand the probe depth of TEY but neglects  
secondary electrons: http://www.icdd.com/resources/axa/vol44/v44_058.pdf


For the K-edges in the hard x-ray regime, with TEY it would not be  
unrealistic to expect probe depths on the order of 1000 angstroms (100  
nm).


Jeff

On Feb 20, 2008, at 6:33 PM, Jenny Cai wrote:


Hi all,

Could someone please answer my questions? I would really appreciate  
your help.


1. For linear combination fitting, there are three indicators for  
the goodness of fitting: R-factor, chi-square and reduced chi- 
square. Could anyone tell me how they work?


2. Since TEY is sensitive for the surface and FY for the bulk (and  
surface?), species detected by TEY should be also detected by FY,  
right?


3. How to calculate the maximum analysis depths for TEY and FY?

Thank you in advance.


Jenny Cai




___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


[Ifeffit] sorting files in Athena

2008-02-21 Thread Zajac, Dariusz
Hi,
I can not find one small thing - sorting files in Athena - it makes me
crazy ;)
I work with QEXAFS data  - usualy few tens of files or even more. Files
are named in such way: file_namerXX.fio where XX is next file number
(1, 2, 3, ... ,10, 11, 12, ... , 20, 21, etc.).
I use Iffefit under Win XP, and every time when I open (import) more
files I have them in such order: 1, 11, 12, 13, ... , 19, 2, 20, 21, 22,
... , 29, 3, 30, 31 ... etc.
So, my question is if there is some hidden way to sort automatically
files during importing or in the menue. Do you know something about? Is
the sorting lexigraphical (from Windows)? 

another problem which apears, is the limitation of the memmory...

thanks for answer and greetings
kicaj

Zajac Dariusz A.Deutsches Elektronen-Synchrotron
-Hasylab
phone:  +49-40-8998-2932Notkestr. 85
fax:+49-40-8998-2787   22607 Hamburg
e-mail:  Germany


___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


Re: [Ifeffit] sorting files in Athena

2008-02-21 Thread Richard Mayes
Check out the section on marking groups in the Athena manual to see if
that meets your needs.  This may not be the answer you're looking for, but
it's a way to sort data by selecting data sets that have a common
expression that matches your search parameters.  The command is under Mark
then mark regex I think.

The Athena manual is found here

http://cars9.uchicago.edu/iffwiki/Documentation

Richard




 Hi,
 I can not find one small thing - sorting files in Athena - it makes me
 crazy ;)
 I work with QEXAFS data  - usualy few tens of files or even more. Files
 are named in such way: file_namerXX.fio where XX is next file number
 (1, 2, 3, ... ,10, 11, 12, ... , 20, 21, etc.).
 I use Iffefit under Win XP, and every time when I open (import) more
 files I have them in such order: 1, 11, 12, 13, ... , 19, 2, 20, 21, 22,
 ... , 29, 3, 30, 31 ... etc.
 So, my question is if there is some hidden way to sort automatically
 files during importing or in the menue. Do you know something about? Is
 the sorting lexigraphical (from Windows)?

 another problem which apears, is the limitation of the memmory...

 thanks for answer and greetings
 kicaj
 
 Zajac Dariusz A.Deutsches Elektronen-Synchrotron
 -Hasylab
 phone:  +49-40-8998-2932Notkestr. 85
 fax:+49-40-8998-2787   22607 Hamburg
 e-mail:  Germany
 

 ___
 Ifeffit mailing list
 Ifeffit@millenia.cars.aps.anl.gov
 http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit



-- 
Richard Mayes

Barnes Group
450/452 Buehler Hall
Department of Chemistry
University of Tennessee
Knoxville, TN 37996

___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


Re: [Ifeffit] problem with feff.inp

2008-02-21 Thread Matt Newville
It looks like the SCF calculation is having a hard time with the new
potential.   The input file you sent only has atoms out to 2.48 Ang
(though perhaps you truncated it when posting???), but the

   SCF 4 0 30 0.2 1

in your feff.inp means to make a self-consistent potential out to 4
Ang.  Without any atoms out that far, I can imagine it would have
problems.I think you'll probably need a larger cluster than a
single shell for a reasonable XANES simulation.

Hope that helps,

--Matt
___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


Re: [Ifeffit] Questions about Athena and XANES

2008-02-21 Thread Matt Newville
Hi Jeff,

Thanks -- you're right.  I was only considering Auger electrons which
are more surface sensitive than the secondary electrons that actually
dominate most TEY measurements.

--Matt
___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov
http://millenia.cars.aps.anl.gov/mailman/listinfo/ifeffit


Re: [Ifeffit] sorting files in Athena

2008-02-21 Thread Ravel, Bruce
 
 Hi,
 I can not find one small thing - sorting files in Athena - it makes me
 crazy ;)
 I work with QEXAFS data  - usualy few tens of files or even more. Files
 are named in such way: file_namerXX.fio where XX is next file number
 (1, 2, 3, ... ,10, 11, 12, ... , 20, 21, etc.).
 I use Iffefit under Win XP, and every time when I open (import) more
 files I have them in such order: 1, 11, 12, 13, ... , 19, 2, 20, 21, 22,
 ... , 29, 3, 30, 31 ... etc.
 So, my question is if there is some hidden way to sort automatically
 files during importing or in the menue. Do you know something about? Is
 the sorting lexigraphical (from Windows)? 

The file selection dialog sorts alphabetically.  

There is no quick-n-easy way to resort the data groups once in
Athena.  The best solution would be to name the files in a way that
alphabetical sorting will be the same as time sorting.  The way to do
that will be to name the first file f01.fio rather than r1.fio.

 another problem which apears, is the limitation of the memmory...

That is a well known, well discussed problem.  Ifeffit is written in
fortran.  Fortran does not have dynamic memory allocation.  The only
practical solution is tonot read so many data files in at a time.

I wish it were different, but it's not.

B


--
 Bruce Ravel  --- [EMAIL PROTECTED]

 National Institute of Standards and Technology
 Synchrotron Methods Group at Brookhaven National Laboratory
 Building 535A
 Upton NY, 11973

 My homepage:http://cars9.uchicago.edu/~ravel
 EXAFS software: http://cars9.uchicago.edu/~ravel/software/exafs/


winmail.dat___
Ifeffit mailing list
Ifeffit@millenia.cars.aps.anl.gov