lysis using Ifeffit
Subject: Re: [Ifeffit] Basic question in normalization and background removal
On Friday, February 08, 2013 01:22:14 PM Bajamundi Cyril wrote:
> My first naïve approach to the approximation of the normalization
> range is to choose a range such that the post-edge line
On Friday, February 08, 2013 01:22:14 PM Bajamundi Cyril wrote:
> My first naïve approach to the approximation of the normalization range is
> to choose a range such that the post-edge line is parallel with pre-edge
> line, this in my novice eyes this seems to be the case between these line
> two
ps.anl.gov] On Behalf Of Ravel, Bruce
Sent: Friday, February 08, 2013 2:17 PM
To: XAFS Analysis using Ifeffit
Subject: Re: [Ifeffit] Basic question in normalization and background removal
Cyril,
My favorite answer is "don't measure data that way". At the very least, add a
few widely
y 08, 2013 2:28 AM
To: ifeffit@millenia.cars.aps.anl.gov
Subject: [Ifeffit] Basic question in normalization and background removal
Hello,
I have here a very basic XANES data processing. I am but a novice in this XANES
analysis.
In Bruce’s lecture “Athena: Data processing
I” posted in the Diam
Dear Cyril,
I think you should read about normalization of XANES (NEXAFS)
spectra at soft X-rays. It is a common problem of normalization with
the short post-edge region. There are few tricks. I think you can
find answer in already published articles (as an example)
Hello,
I have here a very basic XANES data processing. I am but a novice in this XANES
analysis.
In Bruce's lecture "Athena: Data processing
I" posted in the Diamond Light
Source website, he used the Fe_lepidocrocite.000 example to show how background
removal and normalization is done. For