Gents i just got a 53132A - Opt 010
Serial  : 3404A00xxx
Fw Rev  : 3403

An old lady w. a 10811 OCXO in it - HP 1994 Unit - Nice bright display.

I noticed in the Prog. manual (pg 17) that this applies to my counter : 

***** Snip *****

Agilent 53132A Time Interval Delay Arming
-----------------------------------------

Agilent 53131A and Agilent 53132A Counters with a serial number prefix 
below 3646 are identical in their TI arming modes. 
Both only offer Time Interval Delay, where the STOP trigger of a time 
interval measurement can be delayed by a user-specified time.

***** Snip *****


What am i missing ?? , or what has been added in later hw revs ?


The rest of the "Differences" does not seem to apply , 
as i have Fw Rev: 3403  , not 3402 or earlier.



Upgrading fw:
I have access to the 2010 fw proms -  4613 0738

But can i update my "old counter" ?
I seem to loose my Cal , and i can only do the "simple" TI cal @home.

So i'm not sure i want to do that, if i have to send it to a shop for Cal.


TIA
CFO - Denmark

-- 
E-mail:xne...@luna.dyndns.dk

_______________________________________________
time-nuts mailing list -- time-nuts@febo.com
To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts
and follow the instructions there.

Reply via email to