On Montag, 12. November 2018 00:10:49 CET Guido Trentalancia wrote:
> Please note that a vendor bug affects the latest HMO3xxx firmware
> (V6.005) so that the SCPI command SCPI_CMD_GET_DIG_POD_THRESHOLD
> ("POD<p>:THR?") wrongly returns "USER" instead of "USER1" !
> 
> This makes impossible to validate the response, so we just get the
> string (i.e. we use sr_scpi_get_string() instead of
> scope_state_get_array_option()), otherwise the device fails to open...

This is not a bug but following the spec. Query commands return the short form 
for alphanumeric values, and an index of "1" is the same as an omitted index.

Regards,

Stefan

-- 