On Montag, 12. November 2018 00:10:49 CET Guido Trentalancia wrote: > Please note that a vendor bug affects the latest HMO3xxx firmware > (V6.005) so that the SCPI command SCPI_CMD_GET_DIG_POD_THRESHOLD > ("POD<p>:THR?") wrongly returns "USER" instead of "USER1" ! > > This makes impossible to validate the response, so we just get the > string (i.e. we use sr_scpi_get_string() instead of > scope_state_get_array_option()), otherwise the device fails to open...
This is not a bug but following the spec. Query commands return the short form for alphanumeric values, and an index of "1" is the same as an omitted index. Regards, Stefan -- Stefan Brüns / Bergstraße 21 / 52062 Aachen home: +49 241 53809034 mobile: +49 151 50412019
signature.asc
Description: This is a digitally signed message part.
_______________________________________________ sigrok-devel mailing list sigrok-devel@lists.sourceforge.net https://lists.sourceforge.net/lists/listinfo/sigrok-devel