I've seen some comments and questions regarding the performance of the 5680 Rb's.
Yesterday I took delivery of two 5680's. In case anyone is interested, I'm seeing adev numbers in the 2 to 2.5 E-12 range from 80 thru 200 seconds. The unit under test has been running for approx one hour and I let it warm up for 30 minutes before collecting data. The reference is an fts1050 and the data is being collected with an hp 5370B. Not to bad for $40.00 IMHO. Regards Mark Spencer _______________________________________________ time-nuts mailing list -- [email protected] To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.