The fudge value is generally used to compensate for the channel A vs B internal 
offsets.   

Lady Heather works with the TICC and has an "autotune" function (&a) that can 
calculate and set the FUDGE and TIME2 parameters.   You feed a 1PPS signal into 
a "T" adapter and feed the T outputs to the TICC via matched cables.  It 
calculates  the difference between the chA and chB readings and sets the the 
FUDGE1 to the difference (on mine -315 ps)