On 19/02/15 11:00, Emilia C. Arturo (Emily) wrote:
Hello all.

Hello Emilia (Emily),

I'd like to understand what it is I'm looking at when I use Coot's density fit analysis tool. I recognize that there was a post related to this topic on the Coot bb a while ago --the discussion was on how to interpret the red-ness or green-ness of the density fit plot (https://www.mail-archive.com/coot@jiscmail.ac.uk/msg02995.html) <https://www.mail-archive.com/coot@jiscmail.ac.uk/msg02995.html%29--but> --but it doesn't seem the issue was resolved then (i.e. what does 'red' really mean? is it ...bad?).

I failed to fix this for 0.8.1  I will add it to the list now for 0.8.2.

The colours were set for "typical" (~2.1A on the absolute scale).

Now I have more to ask that involves my using phenix-generated FEMs to build in Coot.

So what I've done is the following: I adjust my model in Coot, using a phenix-generated FEM as the map for fitting, then refine with phenix, and using the refined pdb and reflections file, I use phenix to generate a new FEM. Then I repeat. At some point I learned about Coot's density fit analysis tool and took a look at how my model fits. If the map that is selected in the sidebar of Coot is a FEM, then the density fit analysis plot looks mostly green everywhere - fine. If, however, I select as my map in the Coot sidebar the 2Fo-Fc that phenix had generated along with the latest refined model--the one I'm examining with the density fit analysis tool--then Coot's density fit analysis plot looks red (with values ~ < 0.3), with splashes of orange, barely any green or yellow (with values ~ > 0.3), almost everywhere.

:-( yes - this doesn't work for FEM maps.


So these are my questions: What are the units of the density fit values?

electrons/A^3 with volume normalization.

I think FEM maps are normalized (so that the mean is 0 and standard deviation is 1.0) and this is not typical for maps generated from structure factors on the absolute scale - where you'd expect density values smaller than this by a factor of 2 or 3 (or so)).

So the density values from FEM maps are higher and the validation graph greener.

i.e. What is the calculation that's done?

The map is sampled at the atoms and weighted by occupancy.

I'm surprised that the FEM-dependent density fit graphs look so different (i.e. so green) relative to the graphs generated if my map is set to the 2Fo-Fc from the loaded model;

I know less about FEM maps than I should, but I can refer you to Section 7.9.1 of the documentation:

https://www2.mrc-lmb.cam.ac.uk/Personal/pemsley/coot/web/docs/coot.html#Validation-Graphs

After you've rescaled, I'd be interested in your reflections of the differences.

Regards,

Paul.