I just read through the rest of the discussion. He only thing to add is that I
think the flat sample does play a role in the para part of the focusing.
A real focusing experiment includes the light source entrance slit, “curved
diffraction grating” (what we call our sample), and exit slits all
Angel,
Don't correct peak positions. Rather, use the appropriate parameter in
your refinement software (in Fullprof, usually a combination of zero and
sysin, but then again I usually have data from position sensitive
detectors)(if your instrument was properly aligned, zero should refine
to close
Hi all,
I've never used one of these Anton Paar furnaces, but with some of my
instruments (externally heated, not Pt resistive foil heated) I could actually
use the thermocouple as the sample mount. I also had an instrument that had a
thermocouple in contact with the powder sample, which due
Dr. Alexandre (Alex) F. T. Yokochi
Assistant Professor
Department of Chemical Engineering
Oregon State University
Corvallis, OR 97331-2702
Voice: (541) 737-9357
Fax: (541) 737-4600
http://oregonstate.edu/~yokochia
-Original Message-
From: Yokochi, Alexandre [mailto:alexandre.yoko
Hi everyone,
Does anyone remember offhand how to turn on the output that includes
calculated particle size when refining with a symmetry specific size
model in Fullprof? I haven't looked at the program in a few years and
am afraid I can't quite remember...
Alex Y
Hi all,
Does the error maximize at 2-theta = 90 then decrease again? Back when
I was an actual crystallographer and used an INEL system, this was
observed when the capillary position was not coaxial with the detector
(in fact, when it was behind the proper position).
Alex Y
Kurt,
Back when I was doing this I just purchased Y2O3 from a conventional
chemical supplier and annealed it at high temperature (as high as you
can easily get to, usually something like 1150C unless you have really
fancy furnaces and crucibles) for a few days.
Alex Y
Pam and all,
I too have been looking at materials using in house lab data. Using an
Ag rotating anode (I actually used Ag-k beta for simplicity of the lack
of peak splitting) and an MSC curved imaging plate, and very long
exposures we got data we found quite useful. However, I stopped playing
Patrick,
The usual thing is to correct the data using a 1/(sin theta) correction.
Notice that you'll have to create an ESD format file for GSAS and
calculate what your ESDs (or SUs or whatever) in the corrected data is
or you'll get funny statistics. The goal is I/sigma I stays constant -
figure
Igor,
Back when I used to do this a lot, the easiest way was to just start the
occupancy of the two sites at 1 (e.g., 95% + 5%) then give one site a
given variable and the second the opposite sign for the same variable
(e.g., 20.5 and -20.5). I've been out of the XRD game for a few years,
so
Jorge,
I purchased a Rigaku Rapid a few years ago (now superseded by the Spider, I
believe) which I use extensively for single crystal, powder and thin film work.
Unless you are extremely careful, the powder patterns can have somewhat
broadened peaks which makes doing very high quality
SM,
Having had INEL instruments, I would like to point out that you should
carefully determine the angle between the primary beam and the sample
when you make measurements on both standards and samples. This has a
huge effect on line profiles. Peaks, especially very low angle ones,
will have a
Hi Simon,
Is this presentation part of the materials in your web site? If not, I would
really like to look at it.
Alex Y
Dr. Alexandre (Alex) Yokochi
Assistant Professor
Department of Chemical Engineering
Oregon State University
Corvallis, OR 97331-2702
Phone: (541)-737-9357
Fax:
Hi,
One issue to consider is that many programs, when esd's of each datum
are not given explicitly, estimate these as sqrt(I). Further, many
programs weigh the data using these esd's. Therefore, since it is
desired that esd/I remain a constant, if you do multiply your data by a
constant, the
Armel, Irini, and all,
The option to do bond length calculations and BVS calculations, etc,
is not explicitly given in the HTML fullprof manual. It is, however,
described in the appendix included with the more recent releases of Fullprof
(you know, the big ol pdf file called something
Hi all,
I have neutron data (constant wavelength) I have been working on and
the displacement parameters of most atoms turn out to be negative along one
axis (c in hexagonal setting). The material has relatively low absorption
coefficient (mu*r approx .5), and (as expected) when a
Hi everyone,
I have been interested on how well one of my diffractometers would
do with organic samples (i.e., would it be useful for phase matching, or
even SDPD) and am currently collecting a data set on Sucrose (table
variety). I wonder if any of you that has access to the CSD
Hi everyone,
I have some data where I suspect that an atom rather than sitting
exactly on a special position is actually displaced off of it, and should
essentially form a torus around it. I think I overheard that there is the
possibility of using toroidal displacement factors in one of
Sean,
Thanks, I'll give it a try. I am also to the point where I think
writing a quick utility to do it would be trivial, so maybe I'll do it as
soon as instrumentation quits breaking down (I've almost reached steady
state!).
Did you ever get an answer to your GSAS question?
Hi everyone,
I am wondering whether there is an easy to use tool that will take a
SHELX format intensity file and instrumental parameters (or even without
these) and calculate a powder pattern out of that.
I know that sometime ago someone asked exactly the question I am now
Joe,
You'll have fun and learn a lot of strange things playing with powders (I know
I have).
We have both a Siemens (Bruker) D5000 and two INEL systems. Both give
excellent data, etc. The INEL systems, due to the parallel beam optics are way more
sensitive to alignment.
Howdy everyone,
Concerning the data storage formats, Brian Toby wrote that (some time ago)
am not sure that I would use CIF as my internal format, if I were to
create a database, since it is not designed for searching, but I would
certainly want the ability to
22 matches