On 13:16 Wed 11 Nov     , Matthias Zebisch wrote:
> What is the optimal wavelength for Sulfur SAD phasing?
> Is it 1.9A or should one go below that to reduce absorption/damage.

Some additional advice about factors other than wavelength: For S-SAD on 
a home Cu source, my previous lab showed that optimizing (NOT 
maximizing!) redundancy and using *lower* resolution were important. The 
reference:

G. N. Sarma and P. A. Karplus. "In-house sulfur SAD phasing: a case 
study of the effects of data quality and resolution cutoffs." Acta 
Cryst. (2006). D62, 707-716.

-- 
Thanks,
Donnie