On 10/30/12 15:02, Bosch, Juergen wrote:
calculate an anomalous map, you should see the Zn signal even if you collected at the SeMet peak.
Jürgen
......................

Ca can have a noticeable anomalous signal of its own, if your data are good.

If the possibility to collect new data exists, collecting above and below the Zn edge should be informative. (Zn K edge 1.28 A = 9.6586 keV)

--
=======================================================================