On Mon, 28 Sep 2026 17:00:33 +0200 Sabine Schneider <[email protected]> wrote:
> I combined data from two different crystals collected at two > different beamlines, by scaling them together with aimless. However, > the data were collected at slightly different wavelength: 0.968A and > 0.976A and in the output file from aimless it states: 0.97192A. Dear Sabine, not an answer to your question, but I still would correct the AI agent's answer (and stimmulate a bug fix in aimless). In case you combe data collected with different wavelength, and the file format does not support flagging entries per wavelength so that you must decide on reporting only one wavelength, you should always use the shortest wavelength. The reason is simple crystallography: take a crystal that diffracts beyond the limit lamba/2 (rare for MX, but relevant for this question), you might end up with hkl-indices that were impossible to measure. MTZ is quite flexible and I am sure one can include the wavelength per reflection, which would circumvent the problem (and better the PDB adjusts to good practice than the community adjusts to poor implementations). Best, Tim -- Tim Gruene Head of the CF Crystal Structure Analysis Faculty of Chemistry University of Vienna Phone: +43-1-4277-70202 GPG Key ID = A46BEE1A ######################################################################## To unsubscribe from the CCP4BB list, click the following link: https://www.jiscmail.ac.uk/cgi-bin/WA-JISC.exe?SUBED1=CCP4BB&A=1 This message was issued to members of www.jiscmail.ac.uk/CCP4BB, a mailing list hosted by www.jiscmail.ac.uk, terms & conditions are available at https://www.jiscmail.ac.uk/policyandsecurity/