Dave,

On the K2 the way to center the RIT is to remove the knob, turn the shaft until the frequency difference between RIT ON and OFF is zero, then replace the knob with the pointer straight up (with care not to move the shaft when installing the knob).

73,
Don W3FPR

[EMAIL PROTECTED] wrote:
I have a K2 that when I activate the RIT the received frequency drops 600 Hz, too low to tune back to the frequency of the VFO so the contact is lost. The only way that the RIT is on frequency with the VFO A is to turn the RIT on before beginning the contact. Then everything is as it should be….. Any suggestions? I don’t see anything in the manual to explain how to adjust the RIT or center it on the VFO frequency.