Ray, see Amd 1 of EN 50061 for EMC requirements for pacemakers. Section
6.3.4 addresses static fields. A PM must not be affected by 1mT. For less
than 10 mT, the PM must function normally after the field is removed.


Richard Woods

        ----------
        From:  Ray Levasseur [SMTP:[email protected]]
        Sent:  Friday, December 24, 1999 12:23 PM
        To:  [email protected]
        Subject:  Pacemaker magnetic immunity


        Hi Group,
        First I would like to thank everyone who responded to my earlier
conducted 
        emission problem and the valuable suggestions. I have another issue
that has 
        been brought to my attention by one of our customer reps. I have a
customer 
        that is concerned about a couple of their equipment operators that
have 
        heart pacemakers and the effects on their pacemaker by magnetic
fields. Our 
        product has some very strong dc magnets (not a magnetic field
induced by 
        current flow just bar magnets)and they are concerned about the
health of 
        these operators. I would appreciate any information I could pass on
to this 
        concerned customer. Is there a standard for pacemaker magnetic
immunity such 
        as dc magnetic strength, measured magnetic field in A/m or micro
Tesla 
        measured at a defined distance and frequency? I would like to know
if our 
        product has the potential to cause harm to people with a pacemaker.
Any 
        suggestions for testing this would be greatly appreciated. All the
best for 
        Christmas and the New Year.

        Ray Levasseur
        EMC Compliance



        ______________________________________________________
        Get Your Private, Free Email at http://www.hotmail.com


        ---------
        This message is coming from the emc-pstc discussion list.
        To cancel your subscription, send mail to [email protected]
        with the single line: "unsubscribe emc-pstc" (without the
        quotes).  For help, send mail to [email protected],
        [email protected], [email protected], or
        [email protected] (the list administrators).
        

---------
This message is coming from the emc-pstc discussion list.
To cancel your subscription, send mail to [email protected]
with the single line: "unsubscribe emc-pstc" (without the
quotes).  For help, send mail to [email protected],
[email protected], [email protected], or
[email protected] (the list administrators).

Reply via email to