On Sat, Mar 22, 2025 at 01:50:56AM +0800, 自由如风 via Gwyddion-users wrote: > I have obtained .spm and .pfc files from an atomic force microscope,
An atomic force microscope a bit vague, considering that Gwyddion supports 150+ SPM file formats. Pleas see https://sourceforge.net/p/gwyddion/discussion/fileformats/thread/ee82b3156c/ The extension .pfc suggests Bruker Dimension Icon & Nanoscope. So that's what I assume below. > which contain valuable data on modulus, adhesion force, and dissipated > energy. However, I am facing difficulties exporting these specific datasets > (modulus, adhesion force, dissipated energy) to Excel. Notably, I can > successfully export height data, but the mechanical properties > mentioned above do not appear to export correctly. Exporting one channel correctly and another wrong should be pretty much impossible. What is, unfortunately, quite possible is that some data are *imported* wrong to Gwyddion. Meaning the values you see in Gwdddion do match the instrument software. The Nanoscope file format is complicated and has evolved over time. We read many files correctly, but not all. So, do the data really become wrong only after the export? Or are they already wrong in Gwyddion? There is no export to Excel. So I assume you exported the data to plain text files. But really, I have no idea. Regards, Yeti _______________________________________________ Gwyddion-users mailing list Gwyddion-users@lists.sourceforge.net https://lists.sourceforge.net/lists/listinfo/gwyddion-users