The basic Quantum 600 batteries should last 2+ years providing that they
are charged each night after use. The batteries should be charged
overnite 8-10 hours for a full charge.
Now the power recline option can reduce battery charge as well as the
battery life.
To be sure, have each battery load tested to determine holding amps.
Make sure that all your connections are clean and free of energy robbing
corrosion. Plug the charger in and try to use your chair. If the
chair moves, you have a problem with your charger.
If the chair doesn't move, as it shouldn't, then your circuit system is
operating as designed.
Recharge the batteries and measure the total load or volts, which should
read, 28 volts on brand new batteries and 26+ volts on batteries that are 1 year
old. That should help in making a determination if you really need new
batteries.
Best Wishes
W
In a message dated 10/29/2006 10:31:06 A.M. Central Standard Time,
[EMAIL PROTECTED] writes:
My chair is a 1 year old Quantum 600 rehab with power recline & manual foot rest. I charged it for 10 hours. The battery's the original. |