Dear Silvina,

The purpose of divergence slits is to increase the intensities of reflections at high angle, the flip side is an increase in background and peak broadening (that
will extenuate peak overlap) at high angle.

For structural work it is best to stay with fixed slits, no modification needed
for the profile function  - though TOPAS (academic) is the only program
to date (I know of) that can model divergence slits.

Old school techniques of counting for longer at high angle is the best way to
improve stats and will not unduly effect your profile function.

Divergence slits are useful for phase identification especially if you are
processing a bulk number of samples. Stick with fixed slits for Rietveld
analysis.

Regards

William Bisson
CCP14 administrator
http://www.ccp14.ac.uk

Quoting Silvina Pagola <[EMAIL PROTECTED]>:

Hi,
I have a question regarding the use of fixed radiated length data in a laboratory powder diffractometer, with Bragg-Brentano geometry and programmable divergence and antiscatter/receiving slits, which can also be used in the "fixed slit" mode (in which the irradiated length on the sample varies with the theta angle). This is, for what cases the fixed radiated lenght is recommended instead of fixed slits data?. I have only these two choices for structure solution. For one compound I solved, it seems to work better the fixed radiated length data, although the background increases at high angles and I have to refine absorption (surface roughness), to get positive thermal factors. Could someone explain why to use one dataset or the other for structure solution and refinement?
Thanks,
Silvina.



Reply via email to