Prior to SRM660a, we have used annealed powder of pure Y2O3.
This was due to SRM660 having several sub-percent impurities,
which made it unusual for determining the instrumental profile.
Now, we use SRM660a, this is clean.

Regards

Joerg Bergmann

Am Dienstag, den 21.08.2007, 13:12 +0530 schrieb Murugesan S:
> Dear Rietveld useres,
> 
> I need one help, 
> 
> I need to calibrate the machine, and instrumental parameters of our
> INEL diffractometer, 
> 
> I don't have  the LaB6 powder, whether I can use the well grinned,
> annealed CeO2 powder to get the Instrumental Resolution Parameter.  Or
> What type of material suitable for to get the good instrumental
> parameters from the XRD data. 
> 
> Mostly we used it for the size / strain analysis purpose.
> 
> please suggest me the good materials,
> 
> i expect your comments and suggestions
> 
> Thank you,
> 
> 
> with best regards
> SM
> 