Hi Caroline,
This may sound a little crazy,
but I was working on some mummified tissue back in the 80's, (Mommy is old),
and to deal with the dried out tissue we would give them,
1. A quick .5 percent Sodium Hydroxide soak, 5minutes
2. Wash is running water, 1 minute
3. Soak in a 3 parts
Hi,
Does anyone have a chirp proof way of cleaning coplin jars?
Does anyone store special stain dyes and solutions in coplin jars other than
glass?
If so, can you share with me your cleaning protocol and which coplin jars you
use?
We've had too many issues in our lab and washroom with
Hey Banjo,
Yes, I have been inspected without any real representative from histology.
My current employer sends a Pathologist, PhD and the AP Manager
(Cytotechnologist with no real Histology experience) to our CAP inspection.
The person really doesn't need to know anything about Histology
RBS is the same as Methylene Blue, I believe.
_
Jessica Riggleman | Research Associate
Globus Medical, Inc.
Valley Forge Business Center
2560 General Armistead Avenue | Audubon, PA 19403
Ph: (610) 930-1800 ext. 2583 | Fax:
With most microtomes, 1 complete turn of the wheel equals one advance of
four microns, or whatever you have the micrometer set on. What people
don't realize with the HM355 is that is doesn't take a complete turn of the
fine advance wheel to advance the specimen.
If I recall correctly, the
Lauren,
What is your pump made out of?
_
Jessica Riggleman | Research Associate
Globus Medical, Inc.
Valley Forge Business Center
2560 General Armistead Avenue | Audubon, PA 19403
Ph: (610) 930-1800 ext. 2583 | Fax:
We are currently having the same problem with 2 of ours. Stepper motor
replacement did not work either. I would be interested to hear any responses
as well.
Aimee
-Original Message-
From: Sandra Cheasty via Histonet [mailto:histonet@lists.utsouthwestern.edu]
Sent: Friday, September
Hello,
We have 2 Microm HM355S motorized microtomes. We use the motor
to move the block close to the blade, but turn the wheel manually when
sectioning. Both are "jumping" forward on their own occasionally while
sectioning, which is unsettling and chunks into the block. One of
Good morning and happy Friday everyone!
I am looking to buy some siphon drum pumps- does anyone out there use them? We
have three and they are all leaking at the squeeze/bulb area from the xylene
breaking it down. Does anyone know if they manufacture some that are coated to
protect against
Hello Netters,
Does anyone have a protocol for optimizing section quality for specimens with
considerable adipose tissue to produce cryostat sections? We have had several
instances recently with derm specimens with considerable subcutaneous adipose
tissue as well as lymph nodes with
10 matches