Hi all, I have several HP/Agilent frequency counters 53151A (and 53152A) in my lab. Several of them meanwhile show trigger problems.
I have tested the A and B channel between 5 MHz and 200 MHZ at three input levels of -10 dBm, 0 dBm and +10 dBm. Those with trigger problems show erroneous numbers at -10 dBm input level rather selectively, i.e. mostly at 100 MHz, some also at 150 MHz, while the are o.k. Has anyone on this list experienced similar defects and can give a recommendation how to fix/repair them? Best regards Bernd DK1AG www.axtal.com _______________________________________________ time-nuts mailing list -- [email protected] To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.