Recently I have been testing some Quartz oscillators with exceptional stability. (low parts in 10-13th)
I've noticed something that jumped out at these performance levels! Normally I test oscillators on a DMTD system with either an ultra stable Quartz or a Hydrogen Maser reference. On these latest oscillators at the longer Tau (>100sec.) the Quartz versus Quartz data shows much better performance than the Quartz versus Maser! What is happening (I think) in this case is that both Quartz units have exceptionally low and similar ageing. This similar ageing rate masks the true stability performance at longer Tau. The Maser data (with no effective ageing in this case) reveals the true stability at the longer Tau! Cheers! Corby Dawson _______________________________________________ time-nuts mailing list -- [email protected] To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.