On 09/29/2016 11:45 AM, Scott McGrath wrote:
It depends on the beam tube   There is a fixed amount of Cs in any
given beam tube this pool of Cs is consumed during operation.    Once
the Cs is depleted the tube will no longer function.    It's possible
with intermittent use for the tube to run for decades as long as
vacuum maintenance is performed.   Most Cs tubes have a published
expected lifetime before replacement is required  while maintaining
published specifications.   It's possible to extend tube life
somewhat by increasing oven temp and electrode voltages but the
signal becomes noisier as a result

Aha, that's what I was thinking---the manual was misleading and made no sense in this regard. Thanks!

-Ruslan