From: Kent Jones@313C@Eng_Srvc, on 1/27/98 9:48 AM:
This response (from Kent Jones - UL Santa Clara) pertains to UL 3101 and
related standards:
I confirm that EN 61010-1 is available with Amendment 2 to IEC 1010-1
built in (published by BSI and possibly others). I also confirm that
UL1262 is no longer in effect for new Laboratory Equipment designs; UL 3101
applies.
UL 3101-1 excludes "Measurement and Test" equipment which is covered by UL
3111-1 (or it could be evaluated to UL 1244 which is an older
non-harmonized Standard).
Please feel free to write to me directly with further UL 3101 / UL 3111 /
UL 1244 / UL 1262 questions.
Kent Jones
_________________________________________________________________
KENT JONES
Underwriters Laboratories Inc. Voice : (408) 985-2400 x32371
1655 Scott Blvd. Fax: (408) 556-6052
Santa Clara, CA 95050 E-mail: [email protected]
_________________________________________________________________