Hi Marta,
Matt has the right idea here. To get the 1.6-2mm at 7T, I would recommend using
IPAT 2, partial fourier 6/8, TE ~20ms. Use the smallest MB factor that will
give you TR = 1 second (should be MB = 5).
Joseph
On Feb 23, 2015, at 10:01 AM, hcp-users-requ...@humanconnectome.org wrote:
Hi Joseph,
Thanks for your recommendation!
All the best,
___
Marta Moreno-Ortega, Ph.D.
Postdoctoral Research Fellow
Division of Experimental Therapeutics
New York State Psychiatric Institute
Department of Psychiatry
Columbia University College of Physicians and Surgeons
On Feb
Dear HCP users,
I have some questions about the extensively processed rfMRI data of
“parcellation-timeseries-netmats”. Their filenames are
“groupICA_3T_Q1-Q6related468_MSMsulc.tar.gz”,
“NodeTimeseries_3T_Q1-Q6related468_MSMsulc_ICAd*_ts*.tar.gz”, and
Hi
On 25 Feb 2015, at 03:39, Aaron C aaroncr...@outlook.com wrote:
Dear HCP users,
I have some questions about the extensively processed rfMRI data of
“parcellation-timeseries-netmats”. Their filenames are
“groupICA_3T_Q1-Q6related468_MSMsulc.tar.gz”,