Seclaman seclaman_alexan...@yahoo.com
Subject: RE: Configuration of PANalytical X'pert Pro
To: rietveld_l@ill.fr
Date: Monday, November 30, 2009, 2:05 PM
Thank
you for all of your answers, it was more than I hoped for.
I did a Rietveld refinement on stibnite (Sb2S3). And I
mesured it from 5 2theta to 90
/solovyovleonid
***
--- On Mon, 11/30/09, Alexandra Seclaman seclaman_alexan...@yahoo.com wrote:
From: Alexandra Seclaman seclaman_alexan...@yahoo.com
Subject: RE: Configuration of PANalytical X'pert Pro
To: rietveld_l@ill.fr
Date: Monday, November
...@yahoo.com
Subject: RE: Configuration of PANalytical X'pert Pro
To: rietveld_l@ill.fr
Date: Monday, November 30, 2009, 2:05 PM
Thank
you for all of your answers, it was more than I hoped for.
I did a Rietveld refinement on stibnite (Sb2S3). And I
mesured it from 5 2theta to 90. But? from 5 to 20 2Θ
Seclaman
seclaman_alexan...@yahoo.com
a.c.secla...@gmail.com
--- On Sun, 11/29/09, Tony Raftery a.raft...@qut.edu.au wrote:
From: Tony Raftery a.raft...@qut.edu.au
Subject: RE: Configuration of PANalytical X'pert Pro
To: Alexandra Seclaman seclaman_alexan...@yahoo.com
Date: Sunday, November 29, 2009
Dear Alexandra,
The term “standard configuration” is rather indefinite since PANal has a lot of
various configurations and accessories. The most straightforward option for
reducing the low-angle background is the programmable divergence and
anti-scatter slits. For a fixed slit system the