Re: [time-nuts] SR620: cal error 23
On Mon, Mar 18, 2013 at 04:28:32PM +0100, Volker Esper wrote: Hi all, While auto calibrating the SR620 I get the message cal error 23 - the manual isn't that specific about this message, it only states, that error numbers 19 to 23 are caused by Stop TAC problems. The manual also states that error codes 19-23 are the same as the start values (n - 16), so 23 minus 16 should be the same as '7' i.e. 'stop linearity byte out of range' However, the counter counts, but I can't do automatic calibration any more. Anyone having experience with that? Sorry, no further experience with that HTH, Herbert Thanks a lot Volker ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there. ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.
Re: [time-nuts] SR620: cal error 23
Am 18.03.2013 18:00, schrieb Herbert Poetzl: The manual also states that error codes 19-23 are the same as the start values (n - 16), so 23 minus 16 should be the same as '7' i.e. 'stop linearity byte out of range' Where do you read that, I can't find it :-/ Volker ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.
Re: [time-nuts] SR620: cal error 23
On Mon, Mar 18, 2013 at 07:34:26PM +0100, Volker Esper wrote: Am 18.03.2013 18:00, schrieb Herbert Poetzl: The manual also states that error codes 19-23 are the same as the start values (n - 16), so 23 minus 16 should be the same as '7' i.e. 'stop linearity byte out of range' Where do you read that, I can't find it :-/ http://ilrs.gsfc.nasa.gov/docs/timing/sr620_manual.pdf page 40, search for 'stop tac' best, Herbert Volker ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there. ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.
Re: [time-nuts] SR620: cal error 23
Am 18.03.2013 19:53, schrieb Herbert Poetzl: page 40, search for 'stop tac' Herbert, you are the best :-) I didn't expect the error codes in the programming commands section. I'm afraid it could be a hardware failure. Thanks Volker ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.
Re: [time-nuts] SR620: cal error 23
Running Auto Cal a few seconds after switching the counter on, the calibration succeeds, though with values different from the warm ones (who'd have thunk). After warming up, error 23 wins. I have to open it and do the old school trouble shooting with coolant spray... However, any advice, tip, hint would be appreciated. Thank you Volker Am 18.03.2013 20:33, schrieb Volker Esper: Am 18.03.2013 19:53, schrieb Herbert Poetzl: page 40, search for 'stop tac' Herbert, you are the best :-) I didn't expect the error codes in the programming commands section. I'm afraid it could be a hardware failure. Thanks Volker ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there. ___ time-nuts mailing list -- time-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.