Hi Nick, The battery monitoring is multiplexed with the two ADC channels exposed on the external connector (i.e. ADC0 and ADC7). So you cannot sample the battery voltages and these channels at the same time. But you can switch between the two of them quickly if required.
See section 2.4.4. of the Shimmer User Manual for further details. See the FAQ page to see how to convert the raw ADC value you receive back to battery voltage: http://www.shimmer-research.com/links/faqs#S9. Remember to double the value due to the halving of the voltage before it is measured (as mentioned in the referenced section in the manual). Mike On Wed, Apr 11, 2012 at 7:37 AM, Nicholas Hosein <[email protected]>wrote: > I noticed that the shimmer software can access the batter state on the > mote. In my experience the battery monitor and data bus share the same bus > so battery updates cannot be made while collecting data (I could be wrong > about this though). Was there an update to the boiler plate app that allows > battery monitoring (by percentage or voltage) in realtime while collecting > data? > > Much Thanks, > > Nick > > _______________________________________________ > Shimmer-users mailing list > [email protected] > https://lists.eecs.harvard.edu/mailman/listinfo/shimmer-users > >
_______________________________________________ Shimmer-users mailing list [email protected] https://lists.eecs.harvard.edu/mailman/listinfo/shimmer-users