Basically it means the unit has lost the battery backup to the calibration 
memory.   

The cure is a rather simple calibration step.  On the 5372A you input a 60 mV 
10KHz sinewave and perform Test 25.

The adjustment is also called "Maximum Hysteresis"                              
          