The SR620 I worked on was unused for a long time, and would not turn on. Seems the power switch was oxidized and would not self-clean. Had to unmount it so I could use chemical cleaners without getting any residue on the case or PCB. IIRC the whole front panel needed to be pulled to do that (not too difficult). Beware the SR620 is slightly unusual in that the transformer is not switched, but rather the DC supplies are.
Bob L. ________________________________ From: Eric Haskell <[email protected]> To: Time Nuts <[email protected]> Sent: Wed, October 10, 2012 4:59:51 PM Subject: [time-nuts] Stanford RS620 Acting up My RS620 became difficult to start ... _______________________________________________ time-nuts mailing list -- [email protected] To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/time-nuts and follow the instructions there.