In message <ca+oswywkj+lezf_eo424yd-shzpzec8lp6wnx+dsgvw8o+f...@mail.gmail.com>
, Skip Withrow writes:

>You can also see another coil around the beam path, I take it this is
>the C-field coil.  A think the coil around the bottom of the shield is the
>degaussing coil.  The four wires that you see are the connections for these
>two coils.

I would suspect it is the other way around.

The very carefully laid out and mechanically fixed coil is for the
C-field, which should be uniform over the entire microwave field
and orthogonal to beam-path.

The small coil, somewhat haphazardly would in the middle of the
physics package, with a field parallel to the beam-path is much
more likely to be the degaussing coil.

-- 
Poul-Henning Kamp       | UNIX since Zilog Zeus 3.20
[email protected]         | TCP/IP since RFC 956
FreeBSD committer       | BSD since 4.3-tahoe    
Never attribute to malice what can adequately be explained by incompetence.