Dear Yaroslav (and others),
Use cylindrical symmetry for flat plate Bragg-Brentano samples. There
isn't sufficient information in a single scan to do more than that.
Conseqently wnen you used mmm ample symmetry you got "nonsense"
(J=400!!). Use POLFPLOT option "A" to plot the resulting "axial
distributions" (doesn't work for MArch-Dollase).
Bob Von Dreele

R.B. Von Dreele
IPNS Division
Argonne National Laboratory
Argonne, IL 60439-4814



-----Original Message-----
From: Yaroslav Mudryk [mailto:[EMAIL PROTECTED] 
Sent: Friday, February 18, 2005 10:11 AM
To: [email protected]


Hi everyone,

I tried to use spherical harmonics incorporated in EXPGUI/GSAS for 
modelling a preferential orientation in my sample. I used a cylindrical 
sample symmetry and it worked fine but I am confused about using such
type 
of symmetry for a flat sample (Bragg-Brentano goniometer). The rolling 
(mmm) symmetry gave me a texture parameter about 400, which I can not 
accept. I would appreciate any comments related to this topic.

One more question: Does GSAS or any other software have any option for 
visualization of March-Dollase multi-axis preferential orientation?

sorry for using your time,

Yaroslav

Yaroslav Mudryk
252 Spedding Hall
Ames Laboratory
Ames, IA 50011, USA
Phone: 1-(515)-233-2041




Reply via email to