Matt, I tried what you suggested. Unfortunately, it didn't work, msconvert
still told me that it "ignores the wrapper" (see below). I am able to read
the mzXML, but the file contains both positive and negative ion traces. Here
are the command lines and messages from msconvert
--------------------------------------------
C:\Users\cbovet>msconvert
Z:\Toxikologie\LTQVelos\data\Cedric\CBo_1001_110315A\C
Bo_1001_110315A_03.raw --32 --mzXML --filter "analyzerType ITMS" --filter
"polar
ity negative" -o Z:\Toxikologie\MSconvert
format: mzXML (Precision_32 [ 1000514:Precision_32 1000515:Precision_32 ],
ByteO
rder_LittleEndian, Compression_None) indexed="true"
outputPath: Z:\Toxikologie\MSconvert
extension: .mzXML
contactFilename:

filters:
  analyzerType ITMS
  polarity negative

filenames:

Z:/Toxikologie/LTQVelos/data/Cedric/CBo_1001_110315A/CBo_1001_110315A_03.raw

processing file:
Z:/Toxikologie/LTQVelos/data/Cedric/CBo_1001_110315A/CBo_1001_1
10315A_03.raw
[SpectrumListFactory] Ignoring wrapper: analyzerType ITMS
[SpectrumListFactory] Ignoring wrapper: polarity negative
writing output file: Z:/Toxikologie/MSconvert/CBo_1001_110315A_03.mzXML
---------------------------------------------

Do you have another suggestion? Thanks, Cédric


On 29 March 2011 18:58, Matthew Chambers <[email protected]> wrote:

> Interesting. That appears to be a bug in the config file reader. Until it
> gets fixed, use the command-line equivalent for filters, i.e.
> --filter "analyzerType ITMS" --filter "polarity negative"
>
> -Matt
>
>
>
> On 3/29/2011 11:49 AM, Cédric wrote:
>
>> Hi Matt. I used a configuration file (-c) containing the following
>> information
>>
>> outdir=Z:\Toxikologie\MSconvert\negmzXMLfiles
>> mzXML=true
>> 32=true
>> filter="analyzerType ITMS"
>> filter="polarity negative"
>>
>> for the version, it is the one contained with ProteoWizard release
>> 2.1.2596 (ProteoWizard MSdata 2.1.2595, ProteWizardAnalysis 2.1.2587).
>> Is it the information that you need? Thanks for your help
>> Cédric
>>
>> On Mar 29, 5:45 pm, Matthew Chambers<[email protected]>
>> wrote:
>>
>>> What was your exactly command-line and which version of msconvert did you
>>> use?
>>>
>>> -Matt
>>>
>>> On 3/29/2011 3:07 AM, C dric wrote:
>>>
>>>  Hi,
>>>>
>>>
>>>  I am using msconvert to transform my .raw file acquired on a LTQ Velos
>>>> into mzXML file. I performed LC-MS measurements in positive/negative
>>>> mode in a single run, therefore my files contain both ion trace (i.e.,
>>>> positive and negative). I am trying to extract from my .raw file the
>>>> negative ion trace and convert it in mzXML (filter="polarity
>>>> negative"). When I am doing that, msconvert ignore my filter (it wrote
>>>> back Ignoring wrapper: ) and transform my raw data into mzXML file
>>>> which contains the negative and positive ion trace. Does anybody know
>>>> how to extract each ion trace separately?
>>>>
>>>
>>>  thanks a lot for your help, C dric
>>>>
>>>
>>
> --
> You received this message because you are subscribed to the Google Groups
> "spctools-discuss" group.
> To post to this group, send email to [email protected].
> To unsubscribe from this group, send email to
> [email protected].
> For more options, visit this group at
> http://groups.google.com/group/spctools-discuss?hl=en.
>
>

-- 
You received this message because you are subscribed to the Google Groups 
"spctools-discuss" group.
To post to this group, send email to [email protected].
To unsubscribe from this group, send email to 
[email protected].
For more options, visit this group at 
http://groups.google.com/group/spctools-discuss?hl=en.

Reply via email to