Matt, thanks a lot for your help, that's working now! Regards, Cédric

On 30 March 2011 17:33, Matthew Chambers <[email protected]> wrote:

> Overwrite the msconvert in the tpp-bin directory. Or use the full path to
> the new msconvert executable, e.g.
> c:\path\to\msconvert.exe
>
> -Matt
>
>
>
> On 3/30/2011 10:30 AM, Cédric Bovet wrote:
>
>> Ah, that's maybe the point. When I am running msconvert the first time in
>> cmd, it shows me
>> "ProteoWizard release 2.1.2596" at the end. After that, I runned msconvert
>> without args and it shows
>> me ProteoWizard relesae 2.1.2132". What should I do?
>> Cédric
>>
>> Here are the cmd lines that I got
>>
>> On 30 March 2011 16:42, Matthew Chambers <[email protected]
>> <mailto:[email protected]>> wrote:
>>
>>    I can't reproduce this. Are you sure about the version of msconvert
>> you're running?
>>    "ProteoWizard release 2.1.2596" shows up when you run msconvert without
>> args?
>>
>>    -Matt
>>
>>
>>
>>    On 3/30/2011 1:42 AM, Cédric Bovet wrote:
>>
>>        Matt, I tried what you suggested. Unfortunately, it didn't work,
>> msconvert still told me that it
>>        "ignores the wrapper" (see below). I am able to read the mzXML, but
>> the file contains both
>>        positive
>>        and negative ion traces. Here are the command lines and messages
>> from msconvert
>>        --------------------------------------------
>>        C:\Users\cbovet>msconvert
>> Z:\Toxikologie\LTQVelos\data\Cedric\CBo_1001_110315A\C
>>        Bo_1001_110315A_03.raw --32 --mzXML --filter "analyzerType ITMS"
>> --filter "polar
>>        ity negative" -o Z:\Toxikologie\MSconvert
>>        format: mzXML (Precision_32 [ 1000514:Precision_32
>> 1000515:Precision_32 ], ByteO
>>        rder_LittleEndian, Compression_None) indexed="true"
>>        outputPath: Z:\Toxikologie\MSconvert
>>        extension: .mzXML
>>        contactFilename:
>>
>>        filters:
>>           analyzerType ITMS
>>           polarity negative
>>
>>        filenames:
>>
>> Z:/Toxikologie/LTQVelos/data/Cedric/CBo_1001_110315A/CBo_1001_110315A_03.raw
>>
>>        processing file:
>> Z:/Toxikologie/LTQVelos/data/Cedric/CBo_1001_110315A/CBo_1001_1
>>        10315A_03.raw
>>        [SpectrumListFactory] Ignoring wrapper: analyzerType ITMS
>>        [SpectrumListFactory] Ignoring wrapper: polarity negative
>>        writing output file:
>> Z:/Toxikologie/MSconvert/CBo_1001_110315A_03.mzXML
>>        ---------------------------------------------
>>
>>        Do you have another suggestion? Thanks, Cédric
>>
>>
>>        On 29 March 2011 18:58, Matthew Chambers <
>> [email protected]
>>        <mailto:[email protected]>
>>        <mailto:[email protected] <mailto:
>> [email protected]>>> wrote:
>>
>>            Interesting. That appears to be a bug in the config file
>> reader. Until it gets fixed,
>>        use the
>>            command-line equivalent for filters, i.e.
>>            --filter "analyzerType ITMS" --filter "polarity negative"
>>
>>            -Matt
>>
>>
>>
>>            On 3/29/2011 11:49 AM, Cédric wrote:
>>
>>                Hi Matt. I used a configuration file (-c) containing the
>> following
>>                information
>>
>>                outdir=Z:\Toxikologie\MSconvert\negmzXMLfiles
>>                mzXML=true
>>                32=true
>>                filter="analyzerType ITMS"
>>                filter="polarity negative"
>>
>>                for the version, it is the one contained with ProteoWizard
>> release
>>                2.1.2596 (ProteoWizard MSdata 2.1.2595, ProteWizardAnalysis
>> 2.1.2587).
>>                Is it the information that you need? Thanks for your help
>>                Cédric
>>
>>                On Mar 29, 5:45 pm, Matthew Chambers<
>> [email protected]
>>        <mailto:[email protected]>
>>        <mailto:[email protected] <mailto:
>> [email protected]>>>
>>
>>
>>                wrote:
>>
>>                    What was your exactly command-line and which version of
>> msconvert did you use?
>>
>>                    -Matt
>>
>>                    On 3/29/2011 3:07 AM, C dric wrote:
>>
>>                        Hi,
>>
>>
>>                        I am using msconvert to transform my .raw file
>> acquired on a LTQ Velos
>>                        into mzXML file. I performed LC-MS measurements in
>> positive/negative
>>                        mode in a single run, therefore my files contain
>> both ion trace (i.e.,
>>                        positive and negative). I am trying to extract from
>> my .raw file the
>>                        negative ion trace and convert it in mzXML
>> (filter="polarity
>>                        negative"). When I am doing that, msconvert ignore
>> my filter (it wrote
>>                        back Ignoring wrapper: ) and transform my raw data
>> into mzXML file
>>                        which contains the negative and positive ion trace.
>> Does anybody know
>>                        how to extract each ion trace separately?
>>
>>
>>                        thanks a lot for your help, C dric
>>
>>
>>
>>            --
>>            You received this message because you are subscribed to the
>> Google Groups
>>        "spctools-discuss" group.
>>            To post to this group, send email to
>> [email protected]
>>        <mailto:[email protected]>
>>        <mailto:[email protected] <mailto:
>> [email protected]>>.
>>
>>
>>            To unsubscribe from this group, send email to
>>        [email protected]
>>        <mailto:spctools-discuss%[email protected]>
>>        <mailto:spctools-discuss%[email protected]
>>        <mailto:spctools-discuss%[email protected]>>.
>>
>>
>>            For more options, visit this group at
>> http://groups.google.com/group/spctools-discuss?hl=en.
>>
>>
>>        --
>>
>>        You received this message because you are subscribed to the Google
>> Groups "spctools-discuss"
>>        group.
>>        To post to this group, send email to
>> [email protected]
>>        <mailto:[email protected]>.
>>        To unsubscribe from this group, send email to
>> [email protected]
>>        <mailto:spctools-discuss%[email protected]>.
>>        For more options, visit this group at
>> http://groups.google.com/group/spctools-discuss?hl=en.
>>
>>
>>    --
>>    You received this message because you are subscribed to the Google
>> Groups "spctools-discuss" group.
>>    To post to this group, send email to [email protected]
>>    <mailto:[email protected]>.
>>    To unsubscribe from this group, send email to
>> [email protected]
>>    <mailto:spctools-discuss%[email protected]>.
>>    For more options, visit this group at
>> http://groups.google.com/group/spctools-discuss?hl=en.
>>
>>
>> --
>> You received this message because you are subscribed to the Google Groups
>> "spctools-discuss" group.
>> To post to this group, send email to [email protected].
>> To unsubscribe from this group, send email to
>> [email protected].
>> For more options, visit this group at
>> http://groups.google.com/group/spctools-discuss?hl=en.
>>
>
> --
> You received this message because you are subscribed to the Google Groups
> "spctools-discuss" group.
> To post to this group, send email to [email protected].
> To unsubscribe from this group, send email to
> [email protected].
> For more options, visit this group at
> http://groups.google.com/group/spctools-discuss?hl=en.
>
>

-- 
You received this message because you are subscribed to the Google Groups 
"spctools-discuss" group.
To post to this group, send email to [email protected].
To unsubscribe from this group, send email to 
[email protected].
For more options, visit this group at 
http://groups.google.com/group/spctools-discuss?hl=en.

Reply via email to