Hi,
Alan Coelho helped us get our instrument parameter file set up for Topas Academic. We use an X'celerator detector with cobalt radiation (no monochromator) for most of our work.
Here is an example file:
xdd "test.dat"
CoKa7_Holzer(0.001)
Radius(240)
LP_Factor(17)
axial_conv
filament_length 12
sample_length 15
receiving_slit_length 12
primary_soller_angle @ 5Hope this helps.
Laurel Basciano
At 06:48 AM 5/2/2005, you wrote:
Dear colleagues,
Is there anybody who has experience with describing peak shapes using the fundamental parameters approach for a typical X'pert-pro diffractometer setup with an X'celerator detector (fixed incident slits, 0.02 rad sollers, no monochromator)?
Thanks in advance, Arie
******************************************* A. van der Lee Institut Europ�en des Membranes (UMR 5635) Universit� de Montpellier II - cc 047 Place E. Bataillon 34095 Montpellier Cedex 5 FRANCE
visiting address: 300 Av. Prof. E. Jeanbrau
t�l.: 00-33-(0)4.67.14.91.35 FAX.: 00-33-(0)4.67.14.91.19 *******************************************
--
Laurel Basciano
Department of Geological Sciences and Geological Engineering
Queen's University
[EMAIL PROTECTED]
(613)484-1911